 |
In this section a brief
overview is given of the main techniques being used in LCAM-FNWI:
Controled Light Exposure Microscopy
(CLEM)
Fluorescence Fluctuation
Spectroscopy (FFS)
Fluorescence Lifetime Imaging
Microscopy (FLIM)
Fluorescence Recovery After
Photobleaching (FRAP)
Förster Resonance Energy Transfer
(FRET)
Photoactivation Localization
Microscopy (PALM)
Spinning disk microscopy
Total Internal Reflection
Microscopy (TIRF)
Multi-Photon Excitation (MPE)
For more info about microscopy one can visit the
following websites:
Nikon microscopy
education website
Zeiss microscopy education website
Olympus
microscopy education website
also handy:
Invitrogen spectra viewer (spectra of fluorescent dyes)
SIP chart analysis (calibration of your sectioning
fluorescence microscope)
|
CLEM:
Controled Light Exposure Microscopy
Fluorescence microscopy of
living cells enables visualization of the dynamics and interactions of
intracellular molecules. However, fluorescence live-cell imaging is limited
by photobleaching and phototoxicity induced by the excitation light.
Controlled light-exposure microscopy (CLEM),a simple imaging approach that
reduces photobleaching and phototoxicity two- to tenfold, depends on the
fluorophore distribution in the object. By spatially controlling the
lightexposure time, CLEM reduces the excitation-light dose without
compromising image quality.
recommended literature:
Hoebe RA, Van
Oven CH, Gadella Jr TWJ, Dhonukshe PB, Van Noorden CJF, Manders EMM.
Controlled light-exposure microscopy reduces photobleaching and
phototoxicity in fluorescence live-cell imaging, Nat. Biotech. 23,
249 (2005).
Available CAM microscope(s): Nikon C1 and
Nikon A1
|
FFS: Fluorescence
Fluctuation Spectroscopy
Fluorescence
fluctuation techniques like fluorescence correlation spectroscopy (FCS)
and photon counting histogram (PCH) monitor concentrations
and mobility-, binding- and conformational state dynamics of fluorescent
molecules and their complexes in situ. Since FCS and PCH are single-molecule
techniques, molecules f.e. fluorescently labelled proteins can be studied at
the nanomolar level. For many proteins (especially those involved in signal
transduction) this is the physiological relevant concentration in a living
cell, thus no over-expression of the protein is required. For FCS and PCH
the fluorescence intensity is monitored in the small observation volume of a
confocal microscope (green), which is continuously illuminated (blue). A
particle (red) with a given molecular brightness produces an intensity
fluctuation as it passes the observation volume. Particles with a higher
molecular brightness will result in stronger intensity fluctuations. Since
small particles will diffuse more rapidly through the observationvolume than
large molecules, the duration of the fluorescence bursts contains
information on the diffusion speed of the particles.
Both PCH and FCS analysis use the same experimental data, but each technique
focuses on a different property of the signal. While FCS is a measure of the
time-dependent decay of the fluorescence fluctuations yielding parameters
like particle number, diffusion time and dark-state kinetics, PCH calculates
the amplitude distribution of these fluctuations yielding the distribution
of molecular brightness per particle (Chen et al., 1999). When no
fluorescence quenching occurs this distribution provides a direct readout of
the oligomerization state of the particle.
recommended literature:
Schwille & Haustein. Fluorescence Correlation
Spectroscopy: An introduction to its concepts and applications,
www.biophysics.org/education/schwille.pdf
Chen Y, Müller JD, So PT, Gratton E. The photon counting histogram in
fluorescence fluctuation spectroscopy. Biophys J. 77,
553 (1999).
Available CAM microscope(s):
Olympus/Picoquant FLCCS microscope
|
FLIM: Fluorescence Lifetime Imaging
Microscopy
Fluorescence Lifetime Imaging Microscopy (FLIM) is a powerful
tool for producing an image based on the differences in the exponential
decay rate of the fluorescence from a fluorescent sample. It can be used as
an imaging technique. The lifetime of the fluorophore signal, rather than
its intensity, is used to create the image in FLIM. This has the advantage
of minimizing the effect of photon scattering in thick layers of sample.
FLIM is very useful for biomedical tissue imaging, allowing to probe greater
tissue depths than conventional fluorescence microscopy.
recommended literature:
T.W.J. Gadella (Ed.), FRET and FLIM Techniques. Laboratory
techniques in biochemistry and molecular biology 33,
Elsevier Science, Amsterdam (2009)
Available CAM microscope(s):
FLIM (widefield-frequency domain)
and
Olympus/Picoquant FLCCS microscope
(time domain)
|
FRAP: Fluorescence
Recovery After Photobleaching
Fluorescence Recovery After
Photobleaching (FRAP) is a technique to monitor and quantify the
immobility and binding and/or diffusion rate of fluorescent particles. FRAP
is not an equilibrium technique since a selected region will be
photobleached by an intense laser pulse. Hereafter mobile molecules will
enter the photobleached region from the surrounding area and bleached
molecules will move out of the region. The recovery of the fluorescence
intensity contains information about the bindingkinetcs and/or mobility of
the molecules. A related technique is Fluorescence Loss In Photobleaching (FLIP):
After the bleaching pulse fluorescence intensity is monitored in a region
other than the photobleached region. Due to the altered concentration
gradient between bleached area and surroundings, a loss in intensity may be
seen when the two regions are connected. |
recommended literature:
Klonis, N., Rug, M., Harper, I., Wickham, M., Cowman,
A., and Tilley, L. Fluorescence photobleaching analysis for the study of
cellular dynamics. Eur. Biophys. J. 31, 36 (2002).
Sprague, B. L. and McNally, J. G. FRAP analysis of binding: Proper
and fitting. Trends Cell Biol. 15: 84 (2005).
Available CAM microscope(s): all
microscopes
|
FRET: Forster
Resonance Energy Transfer
Fluorescence
Resonance Energy Transfer (FRET) involves the energy transfer through
dipole-dipole coupling of an donor and acceptor chromophore. The resonance
conditions necessary for this process dictate that the fluorescence emission
spectra of the donor overlaps with the absorption spectra of the acceptor
molecule. The degree of overlap is used to calculate the spatial separation,
R, for which energy transfer efficiency, E, is 50% (called the the Förster
radius R0), which typically ranges from 2-7 nm. This range makes
FRET an ideal mechanism for the study of protein-protein interactions and
can be determined by the measurement of fluorescence lifetime, or intensity
of donor or acceptor.
recommended literature:
Jares-Erijman, E.A. and Jovin, T.M. FRET
imaging Nat. Biotech. 21, 1387 (2003)
T.W.J. Gadella (Ed.), FRET and FLIM Techniques. Laboratory
techniques in biochemistry and molecular biology 33,
Elsevier Science, Amsterdam (2009)
Available CAM microscope(s): all
microscopes
|
MPE: Multi-Photon
Excitation
In the
early 1930s two-photon excitation of molecules was predicted by Maria
Göppert-Mayer in the case that high photon densities are present so that two
or more low energy photons can be simultaneously absorbed in a single
quantum event. Since the energy of a photon is inversely proportional to its
wavelength, the two photons should be twice the wavelength necessary for
single-photon excitation. It was until the beginning of the 1990s when Denk
et al. introduced the two-photon laser scanning microscope resulting in the
first biological applications of TPE. Because excitation in multiphoton
microscopy occurs only at the focal point of a diffraction-limited spot, it
is possible to create thin optical sections of thick biological specimens in
order to obtain three-dimensional resolution.
Two-photon microscopy has some major advances over SPE-confocal microscopy
for 3D imaging. First of all is the penetration of near-infrared light used
for TPE much deeper than that of visible light used in conventional (SPE)
confocal microscopy. TPE of thick biological samples allows imaging to a
depth of more than 200 mm whereas SPE confocal microscopy is limited to
depths of approximately 50 mm. In SPE confocal microscopy fluorescent light
is generated throughout the sample along the optical axis but only the
signal from a thin focal plane is detected by placing an aperture (the
so-called pinhole) at the image plane. However, by using TPE molecules are
excited at the focal plane only and therefore no pinhole is required. In
addition, TPE minimizes photobleaching and photodamage in out-of-focus
regions that are usually limiting factors in conventional live cell imaging.
recommended literature:
Diaspro, A., Chirico, G., and Collini, M.
Two-photon fluorescence excitation and related techniques in biological
microscopy. Quart. Rev. Biophys. 38, 97 (2005).
Available CAM microscope(s):
Zeiss LSM510
|
PALM: Photo-Activation Localization
Microscopy
It
is well known that there is a spatial limit to which light can focus:
approximately half of the wavelength of the light you are using. But this is
not a true barrier, because this diffraction limit is only true in the
far-field and localization precision can be increased with many photons and
careful analysis. The image of a point source on a microscope detector is
called the point-spread function (PSF), which is limited by diffraction to
be approximately half the wavelength of the light. But it is possible to
simply fit that PSF with a Gaussian to locate the center of the PSF, and
thus the location of the fluorophore with a much higher accuracy (compare
the 'standard' LSM image <left> with the PALM image <right>).
Betzig et al. (see image from Science)
developed photo-activated localization microscopy (PALM)
while Zhuang and co-workers used a similar technique called stochastic
optical reconstruction microscopy (STORM). In both
techniques samples filled with many dark fluorophores are imaged. The dyes
can be photoactivated into a fluorescing state by a flash of light. Because
photoactivation is stochastic, only a few, well separated molecules "turn
on". Then Gaussians are fit to their PSFs in order to localize the centre of
the particle. After the few bright molecules photobleach (sometimes actively
by using another differently colored excitation source), the next flash of
the photoactivating light activates random fluorophores again and the PSFs
are fit of these different molecules. This process is repeated many times,
building up an image. Because the molecules were switched on-and-off (and
thus localized) at different times, the 'resolution' of the final image can
be much higher than that limited by diffraction. The current limitation
of these techniques is that it can take on the order of hours to collect
enough photons per molecule.
recommended literature:
Betzig E, Patterson GH, Sougrat R, Lindwasser OW, Olenych S,
Bonifacino JS, Davidson MW, Lippincott-Schwartz J, Hess HF. Imaging
intracellular fluorescent proteins at nanometer resolution. Science
313, 1642 (2006).
Rust MJ, Bates M, Zhuang X. Sub-diffraction-limit imaging by stochastic
optical reconstruction microscopy (STORM). Nat Methods 3,
793 (2006).
Available CAM microscope(s):
Andor
Spinning disk microscope
|
Spinning disk microscopy
The
speed limitations imposed by standard confocals can be overcome by
parallelism. An array of beams can be used in parallel either by using
either a line of pinholes or an array of pinholes. The principal of
generating images using an array of pinholes was first proposed in 1883 by
German Physicist Paul Nipkow. The array of scanning pinholes is called a
Nipkow disc after him and they formed the basis of the first television
cameras. The diagram above shows the heart of a practical Nipkow spinning
disc system. A laser is used to illuminate the sample and this is focused
through a pinhole as well to ensure the laser only illuminates the region
and plane of the sample that it is to be imaged. The fluorescence light from
the sample is separated from the illumination by filters. The light from the
sample is re-imaged through the pinhole to reject out of plane light which
will obscure the image and directed towards an imaging system such as an CCD
camera.
recommended literature:
Gräf, R., Rietdorf, J., and Zimmermann, T.
Live Cell Spinning Disk Microscopy. Advances in Biochemical
Engineering/Biotechnology 95, 57 (2005).
Available CAM microscope(s):
Andor Spinning disk microscope and
FLIM
|
TIRF: Total Internal Reflection
Fluorescence
Total
Internal Reflection Fluorescence (TIRF) can be implemented in two ways. Here
‘’objective-based’’ TIRF is implemented. The excitation light that enters
the objective under an angle will hit the interface between the sample glass
and the cell/medium. When this angle is larger than the critical angle,
which is defined by the difference in refractive index between glass and the
sample, the excitation light will be reflected completely. As a consequence
a small evanescent wave is generated in the sample that will excite only
those fluorophores that are in close proximity (~100 nm) of the glass
substrate. TIRF is therefore ideally suited to study membrane processes
without having too much ‘background’-signal from fluorophores located in the
cytoplasm.
recommended
literature:
Schneckenburger, H. Total internal reflection
fluorescence microscopy: technical innovations and novel applications.
Current Opinion in Biotechnology 16, 13 (2005).
Available CAM microscope(s):
|
|
|
|
|